製品名
Cyanide Cell Test, photometric, 0.010-0.500 mg/L (CN-)
usage
sufficient for 25 tests
specific analyte(s)
cyanide
measuring range
0.010-0.500 mg/L (CN-)
technique(s)
photometry: suitable
application(s)
disposal drainage water
drinking water
electroplating
environmental
ground and surface water
mineral water
wastewater
compatibility
for use with Spectroquant® Move 100
for use with Spectroquant® Nova 60 A
for use with Spectroquant® Prove 100 (plus)
for use with Spectroquant® Prove 300 (plus)
for use with Spectroquant® Prove 600 (plus)
detection method
photometric (Barbituric acid, pyridine-carboxylic acid)
storage temp.
15-25°C
Quality Level
General description
This Spectroquant® Cyanide Cell Test allows the accurate quantification of the free and readily liberated cyanide content in aqueous samples.
Method applied: Cyanide ions react with a chlorinating agent to form cyanogen chloride which together with 1,3-dimethylbarbituric acid forms a violet dye (pyridine-free König reaction) that is determined photometrically.
The Spectroquant® Cell Tests come with prefilled 16 mm round cells and all the required reagents to perform the analysis according to the instruction leaflet provided.
All Spectroquant® Cell and Reagent Test Kits are equipped with the unique Live ID (2D barcode) which allows seamless method recognition and contains essential information such as lot number, expiry date, and automatic calibration updates.
Legal Information
signalword
Danger
Hazard Classifications
Aquatic Chronic 3 - Eye Dam. 1 - Met. Corr. 1 - Skin Corr. 1A - STOT SE 3
target_organs
Respiratory system
保管分類
8A - Combustible, corrosive hazardous materials
wgk
WGK 2
試験成績書(COA)
製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。
グローバルトレードアイテム番号
| カタログ番号 | GTIN |
|---|---|
| 1.14561.0001 | 04022536155814 |
ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.