Millipore Sigma Vibrant Logo
Attention: We have moved. Merck Millipore products are no longer available for purchase on MerckMillipore.com.Learn More

118349 Súprava kalibračných štandardov

118349
Purchase on Sigma-Aldrich

Overview

Replacement Information

Products

Catalogue NumberPackaging Qty/Pack
1183490001 4 vials
Description
Catalogue Number118349
OverviewThe Turbiquant® Calibration Standard Set is suitable to calibrate the Turbiquant® 3000 T. The standards offer the following advantages:
1. The new primary standards have approvals which secondary standards do not have (rated to be a primary standard according to US EPA).
2. Instruments used in drinking water analysis no longer need to be certified with separate primary standards.
3. Clear records of final inspection are included in each set.
4. The new standards have an unrivalled precision due to production accuracy, and stability in solution.
5. The standards are ready-to-use.
6. They are non-toxic.
7. They are unproblematic to store and transport.
References
Product Information
HS Code3822 90 00
Applications
Biological Information
Physicochemical Information
Dimensions
Materials Information
Toxicological Information
Safety Information according to GHS
Storage class10 - 13 Iné kvapaliny a tuhé látky
WGKNWG neohrozuje vodné prostredie
Safety Information
Product Usage Statements
Storage and Shipping Information
StorageSkladujte pri +15°C až +25°C.
Packaging Information
Transport Information
Supplemental Information
Specifications
Measured value
- standard 1

0.02 FNU ±0.02 FNU
- standard 210.0 FNU ±1 %
- standard 3100 FNU ±1 %
- standard 41750 FNU ±2 %
Global Trade Item Number
Catalogue Number GTIN
1183490001 04022536902548

Related Products & Applications

Accessories (Primary)

Catalogue Number Description
118336 Prázdne kyvety pre TURBIQUANT® 1500/3000

Related Products By: Brand Facete

Categories

Analytics and Sample Preparation > Test Kits and Photometric Methods > Instrumental Test Systems for Quantitative Analyses > Turbidity Measurements, Turbiquant® > Calibration Standards for Turbidimeter
Analytics and Sample Preparation > Test Kits and Photometric Methods > Base Parameters from A to Z > Turbidity