The selectivity of Chromolith® Prep RP-18 endcapped monolithic columns is comparable to common RP-18 endcapped reversed phase columns. However, the high speed and high efficiency of these preparative columns make them excellent for separations of non-polar basic and acidic compounds as well as peptides.

In most cases, existing methods using particulate columns can be easily transferred to Chromolith® Prep. However, for some applications, it is worth optimizing the method in order to make use of the full potential of this enhanced technology.

Comparison of Chromolith® Prep RP-18 Endcapped 100-25 mm with Particulate Material: Separation of Oxime Derivates

The comparison of Chromolith® Prep RP-18 endcapped with a particulate Purospher RP-18 endcapped column (250-50 mm) shows that both separations have a similar resolution under the same chromatographic conditions. However, Chromolith® Prep RP-18 endcapped has better selectivity than Purospher RP-18 endcapped, as exhibited by the resolution of the additional isomer peak at 7 minutes.

Packing stand NW 50 (250-50) filled with Purospher RP-18 endcapped, 10 μm

Column Packing stand NW 50 (250-50) filled with Purospher® RP-18 endcapped, 10 μm
Mobile Phase A: Water + 0.05% TFA
B: Acetonitrile + 0.05% TFA 
Gradient linear in 30 min up to 100% B 
Flow rate 100 mL/min
Detection UV 210 nm
Sample 125 mg oxime derivates in 500 μL Acetonitrile
Packing stand NW 50 (250-50) filled with Purospher® RP-18 endcapped, 10 μm
Chromolith® Prep RP-18 endcapped 100-25 mm

Column Chromolith® Prep RP-18 endcapped 100-25 mm
Mobile Phase A: Water + 0.05% TFA
B: Acetonitrile + 0.05% TFA
Gradient

linear in 11 min up to 100% B

Flow rate 100 mL/min
Detection UV 210 nm
Sample 125 mg oxime derivates in 500 μL Acetonitrile
Chromolith® Prep RP-18 endcapped 100-25 mm

Separation of Hirudin (Filtrate of Crude Extract)

Without any sample preparation, the crude sample was injected directly onto the Chromolith® Prep RP-18 endcapped column. The separation took only 5 minutes, in which the target product was successfully isolated from the impurities.

Chromolith® Prep RP-18 endcapped 100-25 mm

Column Chromolith® Prep RP-18 endcapped 100-25 mm
Mobile Phase A: Water + 0.1% Formic acid
B: Acetonitrile 100%
Gradient Time [min] %A %B
0 90 10
10 70 30
10.1 90 10
Flow rate 60 mL/min
Detection UV 254 nm
Sample 23 mg hirudin (filtrate of crude extract) injected in 5 mL solution
Separation of Hirudin (filtrate of crude extract)

Separation of Dihydropyridines (Nifedipine, Nimodipine and Nisoldipine)

With the monolithic silica rod technology of Chromolith® Prep RP-18 endcapped columns, it is possible to speed up separations significantly while reducing back pressure.

Chromolith® Prep RP-18 endcapped 100-25 mm

Column Chromolith® Prep RP-18 endcapped 100-25 mm
Mobile Phase A: Water
B: Acetonitrile
Gradient Time [min] %A %B
0 80 20
8 20 80
8.1 80 20
Flow rate 10 mL/min
Detection UV 224 nm
Sample 90 mg mixture of nifedipine, nimodipine and nisoldipine
Separation of Dihydropyridines

Chromolith® Prep Product Range

The ready-to-use columns are cladded with a polymeric material (PEEK), and can be connected directly to any HPLC system.
Product
Ordering Number
Chromolith® Prep Si 125251
Chromolith® Prep RP-18-e 125252
Chromolith® Prep Guard Cartridge Si 125260
Chromolith® Prep Guard Cartridge RP-18e 125261
Chromolith® Prep Sealing Set (2 O-rings) 125254
Chromolith® Prep Tool Set (1 mounting tool filter, 1 mounting tool, 1 hook wrench) 125255
Chromolith® Prep Endcap Set (1 inlet cap complete, 1 outlet cap) 125256
Chromolith® Prep Frit Set (10 frits) 125257
Chromolith® Prep 25 mm Column Couple 125259
 
HPLC Application Collections

HPLC Application Collections

Application Notes and Methods

Learn More
Chromatography Application World

Chromatography
Application World

The ultimate source for the latest news in your field

Learn More