Comparison with Particulate Material
The selectivity of Chromolith® Prep RP-18 endcapped monolithic columns is comparable to common RP-18 endcapped reversed phase columns. However, the high speed and high efficiency of these preparative columns make them excellent for separations of non-polar basic and acidic compounds as well as peptides.
In most cases, existing methods using particulate columns can be easily transferred to Chromolith® Prep. However, for some applications, it is worth optimizing the method in order to make use of the full potential of this enhanced technology.
Comparison of Chromolith® Prep RP-18 Endcapped 100-25 mm with Particulate Material: Separation of Oxime Derivates
The comparison of Chromolith® Prep RP-18 endcapped with a particulate Purospher RP-18 endcapped column (250-50 mm) shows that both separations have a similar resolution under the same chromatographic conditions. However, Chromolith® Prep RP-18 endcapped has better selectivity than Purospher RP-18 endcapped, as exhibited by the resolution of the additional isomer peak at 7 minutes.
Chromolith® Prep RP-18 endcapped 100-25 mm
Column |
Chromolith® Prep RP-18 endcapped 100-25 mm |
Mobile Phase |
A: Water + 0.05% TFA B: Acetonitrile + 0.05% TFA |
Gradient |
linear in 11 min up to 100% B
|
Flow rate |
100 mL/min |
Detection |
UV 210 nm |
Sample |
125 mg oxime derivates in 500 μL Acetonitrile |
|
Separation of Hirudin (Filtrate of Crude Extract)
Without any sample preparation, the crude sample was injected directly onto the Chromolith® Prep RP-18 endcapped column. The separation took only 5 minutes, in which the target product was successfully isolated from the impurities.
Chromolith® Prep RP-18 endcapped 100-25 mm
Column |
Chromolith® Prep RP-18 endcapped 100-25 mm |
Mobile Phase |
A: Water + 0.1% Formic acid B: Acetonitrile 100% |
Gradient |
Time [min] |
%A |
%B |
0 |
90 |
10 |
10 |
70 |
30 |
10.1 |
90 |
10 |
Flow rate |
60 mL/min |
Detection |
UV 254 nm |
Sample |
23 mg hirudin (filtrate of crude extract) injected in 5 mL solution |
|
Separation of Dihydropyridines (Nifedipine, Nimodipine and Nisoldipine)
With the monolithic silica rod technology of Chromolith® Prep RP-18 endcapped columns, it is possible to speed up separations significantly while reducing back pressure.
Chromolith® Prep RP-18 endcapped 100-25 mm
Column |
Chromolith® Prep RP-18 endcapped 100-25 mm |
Mobile Phase |
A: Water B: Acetonitrile |
Gradient |
Time [min] |
%A |
%B |
0 |
80 |
20 |
8 |
20 |
80 |
8.1 |
80 |
20 |
Flow rate |
10 mL/min |
Detection |
UV 224 nm |
Sample |
90 mg mixture of nifedipine, nimodipine and nisoldipine |
|
Chromolith® Prep Product Range
The ready-to-use columns are cladded with a polymeric material (PEEK), and can be connected directly to any HPLC system.
Product
|
Ordering Number
|
Chromolith® Prep Si |
125251 |
Chromolith® Prep RP-18-e |
125252 |
Chromolith® Prep Guard Cartridge Si |
125260 |
Chromolith® Prep Guard Cartridge RP-18e |
125261 |
Chromolith® Prep Sealing Set (2 O-rings) |
125254 |
Chromolith® Prep Tool Set (1 mounting tool filter, 1 mounting tool, 1 hook wrench) |
125255 |
Chromolith® Prep Endcap Set (1 inlet cap complete, 1 outlet cap) |
125256 |
Chromolith® Prep Frit Set (10 frits) |
125257 |
Chromolith® Prep 25 mm Column Couple |
125259 |