Millipore Sigma Vibrant Logo
Attenzione: Ci siamo trasferiti. Non è più possibile acquistare i prodotti Merck Millipore nel sito MerckMillipore.comPer saperne di più

KVVLA2ZTT1 Opticap® XLT 20 with Durapore® - 20" .1 TC/TC with gauge port 1/pk

KVVLA2ZTT1
1  Double Easy-Open bag
Purchase on Sigma-Aldrich

Panoramica

Replacement Information
Description
Catalogue NumberKVVLA2ZTT1
Trade Name
  • Opticap®
DescriptionOpticap® XLT 20 with Durapore® - 20" .1 TC/TC with gauge port 1/pk
References
Product Information
Device ConfigurationCapsule
Connections, Inlet/OutletSanitary Flange/Sanitary Flange
Connections, Vent/Drain1/4 in. Hose Barb with double O-ring Seal; 3/4 in. Gauge Port
Housing MaterialPolypropylene (PP)
Maximum Differential Pressure, bar (psid)Forward: 5.5 bar (80 psid) @ 25 °C; 1.7 bar (25 psid) @ 80 °C. Reverse: 3.5 bar (50 psid) @ 25 °C, intermittent
Maximum Inlet Pressure, bar (psi)5.5 bar (80 psi) @ 25 °C; 2.8 bar (40 psi) @ 60 °C; 1.0 bar (15 psi) @ 80 °C
European Pressure Equipment DirectiveEMD Millipore Corporation® certifies that this product complies with the European Pressure Equipment Directive, 97/23/EC of 29 May 1997. This product has been classified under Article 3 § 3 of the Pressure Vessel Directive. It has been designed and manufactured in accordance with sound engineering practice to ensure safe use. In compliance with Article 3 § 3 of this Pressure Vessel Directive, this product does not bear the CE mark.
Good Manufacturing PracticesThese products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
T-line/Gauge PortT-line with 3/4 in. Gauge Port T-line side
Quality LevelMQ400
Applications
Biological Information
MediaDurapore®
SterilityOther
Sterilization3 autoclave cycles of 60 min @ 126 °C; not in-line steam sterilizable
Bacterial RetentionSamples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
WettabilityHydrophilic
Physicochemical Information
Pore Size0.1 µm
Inlet ConnectionSanitary Flange
Outlet ConnectionSanitary Flange
Air Diffusion at 23 °C≤40 mL/min @ 3.9 bar (56 psig) in water
Bubble Point at 23 °C≥4830 mbar (70 psig) air with water
Gravimetric ExtractablesThe extractables level was equal to or less than 50 mg per capsule after 24 hours in water at controlled room temperature.
Indirect Food AdditiveAll component materials meet the FDA Indirect Food Additive requirements cited in 21 CFR 177-182.
Maximum Inlet Pressure (psig)80 psig
Maximum Inlet Temperature25 °C
Non-Fiber ReleasingThis product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
USP Bacterial EndotoxinsA sample aqueous extraction contains less than 0.5 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) test.
Dimensions
Cartridge Nominal Length20 in. (50 cm)
Device Size20 in.
Inlet Connection Diameter1-1/2 in.
Outlet Connection Diameter1-1/2 in.
Materials Information
Chemistry
  • Polyvinylidene Fluoride (PVDF)
Device Material
  • Polypropylene
Seal MaterialSilicone (SI)
Vent Cap MaterialPolypropylene (PP)
Toxicological Information
Safety Information according to GHS
Safety Information
Product Usage Statements
Storage and Shipping Information
Packaging Information
Material Size1
Material PackageDouble Easy-Open bag
Transport Information
Supplemental Information
Specifications
Global Trade Item Number
Numero di catalogo GTIN
KVVLA2ZTT1 04053252350283