Dot/Slot (Filtration) Blotting

Request Information

Related Resources: Brochures | Application Notes
Detection of transferrin on a dot blot of human serum using Immobilon® Western Chemiluminescent HRP Substrate.
Detection of transferrin on a dot blot of human serum using Immobilon® Western Chemiluminescent HRP Substrate.
Dot blotting is an ideal technique for quickly assessing the levels of a target antigen across many samples at once. Also, it is a popular method for epitope mapping and screening antibodies for target specificity.

Click on the Dot/Slot (Filtration) blotting topics to read about the possible causes and remedies:

Slow or No Filtration of the Sample Through the Membrane

Possible CauseRemedy
Inadequate vacuum
  • Make sure the blotting unit is closed properly and the seal is intact.
  • Make sure the vacuum source (e.g. pump) is operating properly.
  • Seal off any open wells with a high quality laboratory tape. Increase vacuum level.
Membrane pores clogged
  • Centrifuge or filter samples to remove particulates.
  • Dilute viscous samples with buffer.

Back to Top

Little or No Protein Observed on the Blot

Possible CauseRemedy
Not enough protein applied to the membrane
  • Minimize sample dilution and filter more sample through the membrane.
Detergents (e.g., SDS) may inhibit lower molecular weight proteins from binding to the membrane
  • Eliminate detergents if possible.
Stain not sensitive enough.
  • Use a more sensitive stain.

Back to Top

Stained Blot Is Not Uniform

Possible CauseRemedy
Membrane structure was compressed by filter paper
  • Place a second membrane in the blotting unit to protect the membrane receiving the samples.
Air bubbles trapped in the interior of the membrane
  • Pre-wet membrane by laying it on the surface of the methanol. Immersing the membrane can entrap air.
Membrane not pre-wet in methanol
  • Membrane must be pre-wet with methanol; entire membrane should change uniformly from opaque to semi-transparent.
Air bubbles in the sample.
  • Carefully pipette samples into well to avoid the formation of air bubbles.
Not enough sample volume loaded
  • Sample must cover the entire exposed membrane area.

Back to Top

Protein Smeared Across the Top of the Membrane

Possible CauseRemedy
Sample leaked across the wells
  • Make sure the blotting unit is properly assembled, closed and sealed prior to filtration.

Back to Top

Protein Smeared Across the Back of the Membrane

Possible CauseRemedy
Membrane capacity was exceeded
  • Reduce the amount of protein loaded into the well.

Back to Top