Millipore Sigma Vibrant Logo

CTGR73TP1 Aervent CTGR Cartridge Filter 30 in. 0.2 µm Code 7

View Products on Sigmaaldrich.com
CTGR73TP1
1  Double Easy-Open bag
Retrieving price...
Price could not be retrieved
Minimum Quantity is a multiple of
Maximum Quantity is
Upon Order Completion More Information
You Saved ()
 
Request Pricing
Limited Availability
Limited Availability
In Stock 
Discontinued
Limited Quantities Available
Availability to be confirmed
    Remaining : Will advise
      Remaining : Will advise
      Will advise
      Contact Customer Service
      Contact Customer Service

       

      Contact Customer Service

      Overview

      Replacement Information
      Description
      Catalogue NumberCTGR73TP1
      Trade Name
      • Aervent®
      DescriptionAervent CTGR Cartridge Filter 30 in. 0.2 µm Code 7
      References
      Product Information
      HS Code8421 39 90
      Device ConfigurationCartridge
      Maximum Differential Pressure, bar (psid)Forward: 5.5 bar (80 psid) @ 25 °C; 1.8 bar (25 psid) @ 80 °C. Reverse: 4.1 bar (60 psid), intermittent
      Maximum Inlet Pressure, bar (psi)Forward - 5.5 bar (80 psi) @ 25 °C; 1.7 bar (25 psi) @ 80 °C; Reverse - 4.1 bar (60 psi), Intermittent
      Cartridge CodeCode 7
      Quality LevelMQ400
      Applications
      Biological Information
      MediaAervent®
      Sterilization150 (100 forward/50 reverse) SIP cycles of 30 min @ 145 °C
      Bacterial Grow-ThroughRetains Brevundimonas diminuta liquid aerosol for 21 days at a minimum challenge of 10⁷ CFU/cm²
      Bacterial RetentionSamples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
      WettabilityHydrophobic
      Physicochemical Information
      Pore Size0.2 µm
      Bubble Point at 23 °C≥1100 mbar (16 psig) nitrogen with 70/30 % IPA/water
      Gravimetric ExtractablesThe extractables level was equal to or less than 35 mg per 10-inch cartridge after 24 hours in a 70/30% IPA/water mixture at controlled room temperature.
      HydroCorr≤2.3 mL/min @ 2.6 bar (38 psig) per 30 in cartridge
      Nitrogen Diffusion at 23 °C≤72 mL/min @ 970 mbar (14 psig) in 70/30 % IPA/water
      Non-Fiber ReleasingThis product was manufactured with a PTFE membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
      Virus Aerosol RetentionRated for aerosol retention of a total challenge of 108–1010 plaque forming units of ΦX-174 virus particles at 85 Nm3/h (50 SCFM) per cartridge.
      USP Bacterial EndotoxinsAqueous extraction contains <0.25 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) Test
      Dimensions
      Cartridge Nominal Length30 in. (76 cm)
      Diameter6.9 cm (2.7 in.)
      Filtration Area1.95 m²
      Device Size30 in.
      Materials Information
      Chemistry
      • Polytetrafluoroethylene (PTFE)
      Device Material
      • Polypropylene
      Seal MaterialSilicone (SI)
      Support MaterialPolypropylene
      Toxicological Information
      Safety Information according to GHS
      Safety Information
      Product Usage Statements
      Storage and Shipping Information
      Packaging Information
      Material Size1
      Material PackageDouble Easy-Open bag
      Transport Information
      Supplemental Information
      Specifications
      Global Trade Item Number
      Catalogue Number GTIN
      CTGR73TP1 04053252381782