Custom Premix Selecting "Custom Premix" option means that all of the beads you have chosen will be premixed in manufacturing before the kit is sent to you.
Column back pressure at different flow rates with Chromolith® HPLC columns vs. particulate columnsChromolith® HPLC columns owe their rapid separation speed to their unique bimodal pore structure with macro- and mesopores. The macropores reduce the column back pressure and allow the use of faster flow rates, thereby considerably reducing analysis time. Whereas, the mesopores form a fine porous structure, which creates a very large active surface area for high-efficiency separations.
This combination results in separations which are twice as fast as conventional 5 μm particulate columns – at half the back pressure. Furthermore, flow rates can be easily varied from 1 mL up to 9 mL per minute, with the same high-quality resolution. Besides rapid separations, Chromolith® HPLC columns also ensure fast re-equilibration between analyses.
High-Speed Separation of Five Beta-Blockers
A mixture of five beta-blocker drugs demonstrates the extreme time savings and high separation efficiency made possible with Chromolith® columns. Due to the excellent mass transfer properties of the monolithic skeleton, high-speed separation is possible even at a high flow rate. The beta-blockers were well separated with excellent peak symmetry. At 9 mL/min, the analysis time was less than one minute, and the column back pressure was only 153 bar.
Application Example
Column
Chromolith® Performance RP-18 endcapped, 100-4.6 mm
Mobile phase
Isocratic acetonitrile / 0.1% trifluoroacetic acid in water, 20/80 (v/v)