Millipore Sigma Vibrant Logo

MPGL04GK2 Millipak® 40 Gamma Gold Filter

View Products on Sigmaaldrich.com
MPGL04GK2
2  
Price could not be retrieved
Minimum Quantity is a multiple of
Maximum Quantity is
Upon Order Completion More Information
You Saved ()
 
Request Pricing
Limited Availability
Limited Availability
In Stock 
Discontinued
Limited Quantities Available
Availability to be confirmed
    Remaining : Will advise
      Remaining : Will advise
      Will advise
      Contact Customer Service
      Contact Customer Service

       

      Contact Customer Service

      Overview

      Description
      Catalogue NumberMPGL04GK2
      Trade Name
      • Millipak®
      DescriptionMillipak® 40 Gamma Gold Filter
      OverviewThis Millipak® filter delivers particulate- and bacteria-free water at the point of dispense (POD) of water purification systems. Millipak® membrane filters are made with Durapore® PVDF membranes to provide high flow rates and throughputs, low extractable, and broad chemical compatibility. It is gamma-irradiated for additional bacteria control. The filtered water can be used for laboratory work where bacteria control is crucial (cell culture, microbiology media preparation, etc.).
      Product Information
      HS Code8421 21 00
      Device ConfigurationGamma Gold Capsule
      Maximum Differential Pressure, bar (psid)Forward: 4.1 bar @ 25 °C; 1.7 bar @ 80 °C; 0.35 bar @ 123 °C; Reverse: 0.7 bar @ 25 °C
      Maximum Inlet Pressure, bar (psi)5.2 bar @ 25 °C
      Good Manufacturing PracticesThese products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
      Quality LevelMQ300
      Applications
      Application• Sterile 0.22 µm membrane filter for particulate- and bacteria-free water at the point of dispense
      • Use with Alpha-Q, Milli-DI® and Milli-Q® Academic/Biocel/Element/Synthesis/Plus/Gradient systems
      Biological Information
      MediaDurapore®
      SterilitySterile
      Bacterial RetentionSamples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
      WettabilityHydrophilic
      Physicochemical Information
      Pore Size0.22 µm
      Inlet ConnectionNPTM
      Outlet ConnectionStepped Hose Barb with Bell
      Gravimetric ExtractablesThe extractables level was equal to or less than 1.5 mg per sterilized device after a 200 mL flush and 24 hours in water at controlled room temperature.
      Non-Fiber ReleasingThis product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).
      USP Bacterial EndotoxinsA sample aqueous extraction contains less than 0.5 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) test.
      Dimensions
      Inlet Connection Diameter1/4 in.
      Outlet Connection Diameter1/4 in.
      Materials Information
      Chemistry
      • Polyvinylidene Fluoride (PVDF)
      Product Usage Statements
      Usage Statement
      • Directions For Use:
        Organism Retention: Microorganisms
        Mode of Action: Filtering
        Application: General laboratory analysis
        Intended Use: Water purification
        Instructions for Use: This item provides water filtered through 0.22 μm sterilizing-grade membranes. Refer to the system equipment user guide section "Using the System" (available on USB key in the system box).
        Storage Statement: Store in a dry location
        Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
      Packaging Information
      Material Size2
      Global Trade Item Number
      Catalogue Number GTIN
      MPGL04GK2 04053252465260